MANUFACTURER | Oxford Instruments |
---|---|
MODEL | PlasmaLab System 100 – Modular ICP380 |
Samples
- Photomasks sizes : maximum 150 mm (6ˮ); thickness: maximum 6,35 mm (1/4ˮ)
- Samples sizes : maximum 150 mm (6ˮ); thickness: maximum 6,35 mm (1/4ˮ)
Available Temperatures and Gases
- Temperature : -150 °C to 400 °C
- Gases: SF6, C4F8, CF4, O2, H2
Characteristics
- High density 380 mm ICP Source (Generator of 2 MHz, 5 kW)
- 240 mm lower electrode (13.56 MHz, 600 W) with automatic impedance tuning and variable height
- Typical operating pressure: 1 – 100 mTorr
- Fast loading via loadlock
- Etch end point detection system by laser interferometry
- Automated chamber cleaning
- Cryogenic platform
ROUTINE PROCESS
ICP-DRIE Etch
Silicon Etch
- Minimum feature size: 20 nm
- Etch masks: Cr, Al, SiO2 or resist